INSTRUCTIONS FOR SETTING ADJUSTMENTS
The "TS" articulator (47550) is a semi-adjustable,
non-arcon articulator with adjustment settings for
Bennett Angle; condylar inclination; and immediate
sideshift.
You will need to mount magnetic SPLITEX®
plates (46100) onto the articulator prior to
using the unit for the first time. (see separate instructions).
Axis-related models can then be mounted to the articulator
by using a "buffer" plate in-between the
magnetic plates and the quick-setting plaster that
will hold the model.
"TS" adjustment ranges:
Bennett Angle: 0º to 25º
Condylar inclination: 15º to 60º
Immediate sideshift: 0 to 4 mm
Accomplishing settings and movements:
Lateral movement for replication
of excursions can be achieved by loosening the "condyle
set screw" on top of either condylar housing
to allow the condyle balls freedom of movement along
the condylar pathways.
Condylar inclination setting adjustments
can be made by loosening the highest of the two flat
screw-heads on the side of the left and right arms
of the unit; then turning the condylar housing unit
to the appropriate inclination setting, and relocking
the screw-head.
Bennett Angle setting adjustments
can be made by loosening the lower of the two screw-heads
on either arm, lining up the vertical mark for the
appropriate angle-setting with the vertical reference
line, and relocking the screw-head.
Immediate sideshift movement can
be accomplished by loosening both the left and right
set-screw found on the horizontal arm above the condylar
axle, and then loosening the center set-screw. The
threads of the center set-screw will become visible,
with each thread representing 1 mm of lateral movement
(.05 per side).
Revision 0, 3/19/99